Language Reference
SENSe Subsystem

Power Statistics CCDF Measurement

Commands for querying the statistical power measurement of the
complimentary cumulative distribution function (CCDF) measurement
results and for setting to the default values are found in the
Group of Commands" on page
the parameters described in the following commands, are found under
the
key, after the
Meas Setup
selected from the
MEASURE
History:
Added PSTatistic to Basic Mode version A.04.00
Power Statistics CCDF—Channel Bandwidth
[:SENSe]:PSTatistic:BANDwidth|BWIDth <freq>
[:SENSe]:PSTatistic:BANDwidth|BWIDth?
Set the bandwidth that will be used for acquiring the signal.
Factory Preset
and *RST:
5.0 MHz
Range:
10.0 kHz to 6.7 MHz
Default Unit:
Hz
Remarks:
You must be in the Basic, cdma2000, W-CDMA (3GPP),
or W-CDMA (Trial & Arib) mode to use this command.
Use INSTrument:SELect to set the mode.
Power Statistics CCDF—Sample Counts
[:SENSe]:PSTatistic:COUNts <integer>
[:SENSe]:PSTatistic:COUNts?
Set the counts. Measurement stops when the sample counts reach this
value.
Factory Preset
and *RST:
10,000,000
Range:
1,000 to 2,000,000,000
Unit:
counts
Remarks:
You must be in the Basic, cdma2000, W-CDMA (3GPP),
or W-CDMA (Trial & Arib) mode to use this command.
Use INSTrument:SELect to set the mode.