Error Information Query

:SYSTem:ERRor[:NEXT]?
This command queries the earliest entry in the error queue and then
deletes that entry. *CLS clears the entire error queue. It can be used to
continuously monitor the error queue for errors to occur.
Front Panel
Access:
System, Show Errors
Locate SCPI Command Errors
:SYSTem:ERRor:VERBose OFF|ON|0|1
:SYSTem:ERRor:VERBose?
Adds additional information to the error messages returned by the
SYSTem:ERRor? command. It indicates which SCPI command was
executing and where in that command the error was detected.
<error number>,"<error message>;<annotated SCPI command>"
Example:
First set SYST:ERR:VERBOSE ON
If the command SENSe:FREQuently:CENTer 942.6MHz
is sent, then sending SYST:ERR? returns:
−113,"Undefined header;SENSe:FREQuently:<Err>CENTer 942.6MHz $<NL>"
The <Err> shown after FREQuently shows you the
spelling error. (The $<NL> is the typical representation
for the command terminator.
If the command SENSe:FREQuency:CENTer 942.6Sec
is sent, then sending SYST:ERR? returns:
−113,"Invalid suffix;SENSe:FREQuency:CENTer 942.6Sec<Err> $<NL>"
The <Err> shown after Sec shows you the invalid suffix.
Factory Preset
and *RST:
Off. This parameter is persistent, which means that it
retains the setting previously selected, even through a
power cycle.
Remarks:
The verbose SCPI error debugging state is global to all
the SCPI interfaces.
History:
Added version A.04.00