Power Vs. Time (Burst Power) Measurement - Agilent Technologies E4406A VSA Series Programmer's Manual

Transmitter tester
Hide thumbs Also See for E4406A VSA Series:
Table of Contents

Advertisement

Language Reference
SENSe Subsystem

Power vs. Time (Burst Power) Measurement

Commands for querying the power versus time measurement results
and for setting to the default values are found in the
of Commands" on page
parameters described in the following commands, are found under the
key, after the
Meas Setup
from the
key menu.
MEASURE
Power vs. Time—Number of Bursts Averaged
[:SENSe]:PVTime:AVERage:COUNt <integer>
[:SENSe]:PVTime:AVERage:COUNt?
Set the number of bursts that will be averaged. After the specified
number of bursts (average counts), the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST:
10
Range:
1 to 10,000
Remarks:
You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
Power vs. Time—Averaging State
[:SENSe]:PVTime:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:PVTime:AVERage[:STATe]?
Turn averaging on or off.
Factory Preset
and *RST:
Off
Remarks:
You must be in the EDGE(w/GSM), GSM or Service
mode to use this command. Use INSTrument:SELect to
set the mode.
312
233. The equivalent front panel keys for the
measurement has been selected
Power vs Time
"MEASure Group