Channel Power Measurement - Agilent Technologies E4406A VSA Series Programmer's Manual

Transmitter tester
Hide thumbs Also See for E4406A VSA Series:
Table of Contents

Advertisement

Channel Power Measurement

Commands for querying the channel power measurement results and
for setting to the default values are found in the
Commands" on page
parameters described in the following commands, are found under the
key, after the
Meas Setup
from the
key menu. CHPower used instead of the more
MEASURE
std-compliant CPOWer, as that syntax was already used for Carrier
Power measurement (but has since been renamed).
Channel Power—Average Count
[:SENSe]:CHPower:AVERage:COUNt <integer>
[:SENSe]:CHPower:AVERage:COUNt?
Set the number of data acquisitions that will be averaged. After the
specified number of average counts, the averaging mode (terminal
control) setting determines the averaging action.
Factory Preset
and *RST:
20
Range:
1 to 10,000
Remarks:
You must be in the cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or Basic mode to use
this command. Use INSTrument:SELect to set the
mode.
Channel Power—Averaging State
[:SENSe]:CHPower:AVERage[:STATe] OFF|ON|0|1
[:SENSe]:CHPower:AVERage[:STATe]?
Turn averaging on or off.
Factory Preset
and *RST:
On
Remarks:
You must be in the cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or Basic mode to use
this command. Use INSTrument:SELect to set the
mode.
Chapter 5
233. The equivalent front panel keys for the
measurement has been selected
Channel Power
Language Reference
SENSe Subsystem
"MEASure Group of
299

Advertisement

Table of Contents
loading

Table of Contents