Sense]:Acp:offset:list:points <Integer>,<Integer>,<Integer>,<Integer>,<Integer - Agilent Technologies E4406A VSA Series Programmer's Manual

Transmitter tester
Hide thumbs Also See for E4406A VSA Series:
Table of Contents

Advertisement

Range:
0 Hz to 20 MHz for iDEN, Basic
0 Hz to 45 MHz for cdmaOne
10 Hz to 45 MHz for cdmaOne
0 Hz to 100 MHz for cdma2000, W-CDMA (3GPP),
W-CDMA (Trial & Arib)
Default Unit:
Hz
Remarks:
You must be in Basic, cdmaOne, cdma2000, W-CDMA
(3GPP), W-CDMA (Trial & Arib), or iDEN mode to use
this command. Use INSTrument:SELect to set the
mode.
Adjacent Channel Power—Number of Measured Points
[:SENSe]:ACP:OFFSet:LIST:POINts
<integer>,<integer>,<integer>,<integer>,<integer>
[:SENSe]:ACP:OFFSet:LIST:POINts?
Selects the number of data points. The automatic mode chooses the
optimum number of points for the fastest measurement time with
acceptable repeatability. The minimum number of points that could be
used is determined by the sweep time and the sampling rate. You can
increase the length of the measured time record (capture more of the
burst) by increasing the number of points, but the measurement will
take longer. Use [:SENSe]:ACP:POINts to set the number of points
used for measuring the reference channel.
Factory Preset
and *RST:
Mode
Offset A
Basic &
1024
cdmaOne
Range:
64 to 65536
Remarks:
The fastest measurement times are obtained when the
number of points measured is 2
You must be in Basic, cdmaOne mode to use this
command. Use INSTrument:SELect to set the mode.
Chapter 5
Offset B
Offset C
1024
1024
Language Reference
SENSe Subsystem
Offset D
Offset E
1024
1024
n
.
273