Sense]:Spectrum:frequency:span <Freq - Agilent Technologies E4406A VSA Series Programmer's Manual

Transmitter tester
Hide thumbs Also See for E4406A VSA Series:
Table of Contents

Advertisement

Spectrum—Frequency Span
[:SENSe]:SPECtrum:FREQuency:SPAN <freq>
[:SENSe]:SPECtrum:FREQuency:SPAN?
Set the frequency span to be measured.
Factory Preset
and *RST:
1.0 MHz
100.0 kHz for iDEN mode
Range:
10 Hz to 10.0 MHz (15 MHz when Service mode is
selected)
Default Unit:
Hz
Remarks:
The actual measured span will generally be slightly
wider due to the finite resolution of the FFT.
To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Spectrum—Sweep (Acquisition) Time
[:SENSe]:SPECtrum:SWEep:TIME <time>
[:SENSe]:SPECtrum:SWEep:TIME?
Set the sweep (measurement acquisition) time. It is used to specify the
length of the time capture record. If the specified value is less than the
capture time required for the specified span and resolution bandwidth,
the value is ignored. The value is set at its auto value when auto is
selected. This is an advanced control that normally does not need to be
changed.
Factory Preset
188.0 µs
and *RST:
15.059 ms, for iDEN mode
Range:
100 ns to 10 s
Default Unit:
seconds
Remarks:
NOTE: You must be in the Service mode to use this
command. Use INSTrument:SELect to set the mode.
This command only effects the RF envelope trace.
Chapter 5
Language Reference
SENSe Subsystem
335