Power Statistics Ccdf Measurement - Agilent Technologies E4406A VSA Series Programmer's Manual

Transmitter tester
Hide thumbs Also See for E4406A VSA Series:
Table of Contents

Advertisement

Language Reference
MEASure Group of Commands

Power Statistics CCDF Measurement

This is a statistical power measurement of the complimentary
cumulative distribution function (CCDF). You must be in the Basic,
cdma2000, W-CDMA (3GPP), or W-CDMA (Trial & Arib) mode to use
these commands. Use INSTrument:SELect to set the mode.
The general functionality of CONFigure, FETCh, MEASure, and READ
are described at the beginning of this section. See the SENSe:PSTat
commands for more measurement related commands.
:CONFigure:PSTatistic
:FETCh:PSTatistic[n]?
:READ:PSTatastic[n]?
:MEASure:PSTatastic[n]?
History:
Front Panel
Access:
Measurement Results Available
n
0
Returns unprocessed I/Q trace data, as a series of comma-separated trace
points, in volts. The I values are listed first in each pair, using the 0 through
even-indexed values. The Q values are the odd-indexed values,
not specified or
Returns 10 comma-separated scalar results:
n=1
1. Average input power (in dBm)
2. Probability at the average input power level (in %)
3. Power level that has 10% of the power
4. Power level that has 1% of the power
5. Power level that has 0.1% of the power
6. Power level that has 0.01% of the power
7. Power level that has 0.001% of the power
8. Power level that has 0.0001% of the power
9. Peak power (in dB)
10. Count
246
Version A.03.00 or later, added in Basic A.04.00
Measure, Power Stat CCDF
After the measurement is selected, press
to restore factory defaults.
Defaults
Restore Meas
Chapter 5

Advertisement

Table of Contents
loading

Table of Contents