Language Reference
SENSe Subsystem
Waveform—Resolution BW Filter Type
[:SENSe]:WAVeform:BANDwidth|BWIDth[:RESolution]:TYPE
FLATtop|GAUSsian
[:SENSe]:WAVeform:BANDwidth|BWIDth[:RESolution]:TYPE?
Select the type of Resolution BW filter that is used. This is an Advanced
control that normally does not need to be changed.
Flat top - a filter with a flat amplitude response, which provides the
best amplitude accuracy.
Gaussian - a filter with Gaussian characteristics, which provides the
best pulse response.
Factory Preset
and *RST:
Gaussian
Remarks:
To use this command, the appropriate mode should be
selected with INSTrument:SELect.
Waveform—Decimation of Waveform Display
[:SENSe]:WAVeform:DECimate[:FACTor] <integer>
[:SENSe]:WAVeform:DECimate[:FACTor]?
Set the amount of data decimation done on the IQ data stream. For
example, if 4 is selected, three out of every four data points will be
thrown away. So every 4th data point will be kept.
Factory Preset
and *RST:
1
Range:
1 to 4
Remarks:
To use this command, the appropriate mode should be
selected with INSTrument:SELect.