Rf Port Power Range Auto - Agilent Technologies E4406A VSA Series Programmer's Manual

Transmitter tester
Hide thumbs Also See for E4406A VSA Series:
Table of Contents

Advertisement

Language Reference
SENSe Subsystem

RF Port Power Range Auto

[:SENSe]:POWer[:RF]:RANGe:AUTO OFF|ON|0|1
[:SENSe]:POWer[:RF]:RANGe:AUTO?
Select the RF port power range to be set either automatically or
manually.
On - power range is automatically set as determined by the actual
measured power level at the start of a measurement.
Off - power range is manually set
Factory Preset
and *RST:
On
Remarks:
You must be in the cdmaOne, EDGE(w/GSM), GSM,
NADC, PDC, cdma2000, W-CDMA (3GPP), or W-CDMA
(Trial & Arib) mode to use this command. Use
INSTrument:SELect to set the mode.
Front Panel
Access:
Input, Max Total Pwr (at UUT)
308
Chapter 5

Advertisement

Table of Contents
loading