Language Reference
STATus Subsystem
Questionable Calibration Negative Transition
:STATus:QUEStionable:CALibration:NTRansition <number>
:STATus:QUEStionable:CALibration:NTRansition?
This command determines what bits in the Questionable Calibration
Condition register will set the corresponding bit in the Questionable
Calibration Event register when the condition register bit has a
negative transition (1 to 0). The variable <number> is the sum of the
decimal values of the bits that you want to enable.
Factory Preset
and *RST:
Range:
Questionable Calibration Positive Transition
:STATus:QUEStionable:CALibration:PTRansition <number>
:STATus:QUEStionable:CALibration:PTRansition?
This command determines what bits in the Questionable Calibration
Condition register will set the corresponding bit in the Questionable
Calibration Event register when the condition register bit has a
positive transition (0 to 1). The variable <number> is the sum of the
decimal values of the bits that you want to enable.
Factory Preset
and *RST:
Range:

Questionable Frequency Register

Questionable Frequency Condition
:STATus:QUEStionable:FREQuency:CONDition?
This query returns the decimal value of the sum of the bits in the
Questionable Frequency Condition register.
The data in this register is continuously updated and reflects the
NOTE
current conditions.
354
0 to 32767