Solving Problems

If there are network connection problems

See the "Network Troubleshooting Guide" topic in the Agilent Logic
Analyzer application's online help. Briefly:
• The Windows Firewall must be set to allow access to the "Agilent Logic
• The Windows Firewall must be set to respond to the following ICMP
• The link activity light must be on; If these LEDs are not on, the LAN
• If connecting two logic analysis systems directly together (that is, no
If your IP address is in the same subnet as a working computer's (use
"ipconfig /all | more" to view IP addresses on machine in question), it is
possible that your network administrators have denied access to any
resources if the computer is not in the domain. In this case, you need to
add the computer (logic analysis system) to the domain.
132
Analysis" service in order to use remote and multiframe connections
(see
"To restore the logic analysis system firewall defaults"
page 109).
(Internet Control Message Protocol) requests in order to use remote and
multiframe connections:
Allow incoming echo requests.
Allow outgoing destination unreachable.
Allow outgoing time exceeded.
(See
"To restore the logic analysis system firewall defaults"
page 109.)
segment may be dead.
hub or switch between them), make sure you are using a special
cross- over cable.
16900 Series Logic Analysis Systems Installation Guide
on
on