Advertisement

5
Setting Up Multiframe Configurations
9 If you are not able to see the multiframe system in the Overview
10 Power up the performance PC, log in, start the Agilent Logic Analyzer
11 If you have trouble connecting to the frame, check these potential
96
window on the system, do the following:
a Exit the Agilent Logic Analyzer application.
b Power down all systems.
c Disconnect and reconnect all cables between all systems.
d Following the startup sequence described in Step 7 above, power up
all systems.
e Log in to one of the systems to again pretest the multiframe setup,
wait for the Agilent Logic Analyzer application to start again. If it
starts properly, then exit the application.
application, and connect to one of the systems remotely:
a When the application is starting and the Offline Startup Options
dialog appears, click "Go Online...".
b In the Select System to Use dialog, click "Add System".
c In the Frame Host Name or IP Address dialog, enter the name of the
frame, and click "OK".
d Select the system that was just added and click "Details...".
e In the Frame/Module Information dialog, click "Set as Auto- Connect";
then, click "OK".
Now, each time you start the application on the performance PC, it will
auto- connect to that frame of the multiframe logic analysis system.
problems:
a Verify that the multiframe logic analysis system appears in the
Overview window when running the Agilent Logic Analyzer
16900 Series Logic Analysis Systems Installation Guide