Installing Logic Analyzer Cards and Probes
6 To verify the flex cables are installed correctly, use an ohm meter to
7 While maintaining alignment with the cards as a set, insert them the
42
check continuity between the pair of test points associated with each
connector used.
If there is continuity (0 ohms), the flex cable is seated correctly in the
connector.
If there is no continuity (infinite ohms), the flex cable is not seated
correctly. Repeat the flex cable attachment steps, and check the
continuity again.
rest of the way into the slots, seat them, lock the ejector tabs, and
tighten the thumb screws.
16900 Series Logic Analysis Systems Installation Guide