Up-Down Button Board (Buttons Frame Ui Assy) Replacement - GE Vivid E80 Service Manual

Hide thumbs Also See for Vivid E80:
Table of Contents

Advertisement

Remove the Palm Rest ASSY
Install the Palm Rest ASSY

Up-Down Button Board (Buttons Frame UI Assy) replacement

Read and follow
WARNING
Vivid E80/E90/E95
GC091052 Rev. 3
7. Remove Handle Left Top and Handle Right Top.
Follow this procedure to remove the Palm Rest ASSY:
1. Release the lock mechanisms securing the Palm Rest
ASSY.
2. Remove the Palm Rest ASSY.
Follow this procedure to install the Palm Rest ASSY:
1. Ensure the XYZ Buttons cable runs along the lower tray
cable channel so that the cable is not pinched when the
Palm Rest is replaced.
2. Replace the Palm Rest ASSY.
3. Install Handle Left Top and Handle Right Top.
4. Install the lock mechanisms securing the Palm Rest Cover.
5. Replace the Lower Operating Panel.
6. Replace the Upper Operating Panel.
Energy Control and Power Lockout for Vivid E80/E90/E95.
When servicing parts of the Ultrasound system where there is
exposure to voltage greater than 30 volts:
1. Follow LOCK OUT/TAG OUT procedures.
2. Turn off the breaker.
3. Unplug the Ultrasound system.
4. Maintain control of the Ultrasound system power plug.
5. Wait for at least 30 seconds for capacitors to discharge as
there are no test points to verify isolation.
Ultrasound System components may be energized.
Service Manual
Other Top Console Parts replacement
8-59