GE Vivid E80 Service Manual page 325

Hide thumbs Also See for Vivid E80:
Table of Contents

Advertisement

Preparations (continued)
Hard Disk Drive (HDD) removal procedure
Vivid E80/E90/E95
GC091052 Rev. 3
Back End Processor (BEP) parts replacement
When preparing for the replacement, you must perform the
following steps:
1. Record the Ultrasound system's TCPIP settings and
installed Option strings.
2. Back up the Patient Archive images, Report Archive,
User-Defined Configuration (Customer Presets), and
Service settings. You will perform a Restore after the install.
3. Reload the system software from removable media. Select
No when asked: Keep User Data and Patient Archive?
4. Power down the Ultrasound system.
5. Disconnect the Mains Power Cable from the wall outlet.
6. Disconnect all external cables (probes, ECG, physio,
Ethernet, audio, video, USB).
7. Remove the Left Side Cover.
8. Remove the BEP Cover.
Follow these steps to remove the HDD:
1. Disconnect the cables connecting the HDD.
2. Remove the two screws that secures the HDD Bracket to
the BEP's frame.
3. Get hold of the HDD assembly and pull it out.
Service Manual
8-103

Advertisement