Electrical requirements for the Vivid E80/E90/E95
VOLTAGE
100-240 VAC
Site circuit breaker
CAUTION
Site power outlets
Vivid E80/E90/E95
GC091052 Rev. 3
General Ultrasound system requirements
In the table below, the electrical specifications for the Vivid E80/
E90/E95 includes monitor and on-board peripherals.
The specifications apply for the following models:
GC000220 - Vivid E95
GC000230 - Vivid E90
GC000240 - Vivid E80
Table 2-3:
Electrical specifications
TOLERANCES
±10%
The current drain will vary depending on the mains voltage.
At 230 VAC the current may be up to 3 A.
At 100 VAC the current may be up to 7 A.
Power outage may occur. The Vivid E80/E90/E95 requires a
dedicated single branch circuit. To avoid circuit overload and
possible loss of critical care equipment, make sure you do not
have any other equipment operating on the same circuit.
It is recommended that the branch circuit breaker for the
Ultrasound system be readily accessible.
A dedicated AC power outlet must be within reach of the
Ultrasound system without extension cords. Other outlets
adequate for the external peripherals, medical and test
equipment needed to support this Ultrasound system must also
be present within 1 m (3.2 ft.) of the Ultrasound system.
Electrical installation must meet all current local, state, and
national electrical codes.
Service Manual
POWER
CONSUMPTION
700 W / 770 VA
FREQUENCY
50/60 Hz
2-5