Loading the software from Usb Flash Drive (UFD)
–
Vivid E80/E90/E95
GC091052 Rev. 3
Loading the software
1. Power down the Ultrasound system.
2. Remove all external USB devices.
3. Install the UFD in one of the USB ports on the Ultrasound
system.
4. Power on the Ultrasound system and at the same time
press and hold the F11 key on the alphanumeric keyboard
until you get the first menu on the screen.
5. Follow the instructions on the screen.
Service Manual
Loading the software
8-7