Adjust the 2D mode controls
WARNING
Vivid E80/E90/E95
GC091052 Rev. 3
Always use the minimum power required to obtain acceptable
images in accordance with applicable guidelines and policies.
The following controls can be adjusted to optimize the 2D Mode
display:
Swipe to page 2 on the Touch panel and press either Soft or
Sharp Auto Tissue setting.
Soft: optimizes the radial and lateral uniformity and
brightness of the tissue continuously in real-time.
The mention "Soft" is displayed on the upper right
corner of the image area
Sharp: further enhances the image display by
optimizing the grayscale curve.
The mention "Sharp" is displayed on the upper right
corner of the image area
The Auto Tissue setting (Soft or Sharp) can be turned on/off
by pressing Auto on the Control panel. The last used setting
is then applied.
The Auto Tissue settings are only available in live scanning
and cannot be turned off when the image is stored.
If available, press Virtual Apex (probe dependent) to
improve near field imaging, allowing increased visibility up
to the width of the full probe aperture close to the surface.
Use the Gain and TGC controls to optimize the overall
image.
Gain increases or decreases the amount of echo
information displayed. TGC compensates for depth-related
attenuation in the image.
Service Manual