Probe/Connectors Checks - GE Vivid E80 Service Manual

Hide thumbs Also See for Vivid E80:
Table of Contents

Advertisement

Probe/Connectors Checks

NOTE:
CAUTION
Step
Task
1.
Press Probe on the Operating
Panel.
If not already selected, use the
2.
trackball to select the desired
probe.
3.
Trackball to the desired
application.
Press Select to launch the
application.
To change application without
changing the current probe,
press Appl. on the Operating
Panel.
4.
Verify no missing channels.
5.
Verify there's no EMI/RFI or
artifacts specific to the probe.
6.
Test the probe in each active
connector slot.
7.
Do a leakage test on the probe.
Repeat this procedure for all
8.
available probes.
Vivid E80/E90/E95
GC091052 Rev. 3
Probes can be connected at any time, whether the Ultrasound
system is ON or OFF
Take the following precautions with the probe cables:
Keep away from the wheels.
Do not bend.
Do not cross cables between probes.
Table 4-1:
Probe and Connectors Checks
Service Manual
Functional checks
Expected Result(s)
A list of the connected probes will
pop up on the screen.
An application menu for the desired
probe is listed on the screen.
The selected application starts.
All channels are functioning.
No EMI/RFI or artifacts.
It will display pictorial data each
time.
It passes the test.
4-29