Replacement Procedures

Wagon AN Drawer Sheet Met. Assembly replacement

Preparations
8-48
When preparing for the replacement, you must perform the
following steps:
1. Power down the Ultrasound system.
2. Disconnect the Mains Power Cable from the wall outlet.
3. Disconnect all probes and external I/O cabling.
4. Remove the Rotary buttons below the Touch screen and the
Volume button.
5. Remove the Operating Panel, Upper.
6. Remove the Operating Panel, Lower. Place it on a clean
surface with the front down.
7. Remove the A/N Keyboard ASSY.
Vivid E80/E90/E95
GC091052 Rev. 3