Replacement Procedures
Remove the XY Mechanism
Install the XY Mechanism
NOTE:
8-64
3. Disconnect all probes and external I/O cabling.
4. Remove the LCD Rear Cover.
5. Remove the LCD Monitor assembly.
6. Remove the LCD Arm assembly.
7. Remove the Rotary buttons below the Touch screen and the
Volume button.
8. Remove the Operating Panel, Upper.
9. Remove the Operating Panel, Lower. Place it on a clean
surface with the front down.
10. Remove the Bulkhead Board.
11. Remove the Bulkhead Plate.
12. Remove the Upper Frame Assembly.
13. Remove the Lower Frame Assembly.
Follow this procedure to remove the XY Mechanism:
1. From the rear side of the Ultrasound system, remove the
four (4x) fixing screws.
2. Remove the XY mechanism.
Follow this procedure to install the XY Mechanism:
1. Position the XY mechanism so it aligns with the holes for the
four fixing screws.
2. Install the four fixing screws.
3. Install the Lower Frame Assembly.
Ensure that Bumper Z enters the slide at the back, before
tighten the screws.
4. Install the Upper Frame Assembly.
5. Install the Bulkhead Plate.
6. Install the Bulkhead Board.
7. Install the Operating Panel, Lower.
8. Install the Operating Panel, Upper.
9. Install the LCD Arm assembly.
10. Install the LCD Monitor assembly.
11. Install the LCD Rear Cover.
Vivid E80/E90/E95
GC091052 Rev. 3
Service Manual