Diagnostics/Troubleshooting

Motor controller test

Introduction
Test the motor controller
7-10
This test program tests the XYZ Motor Controller. It also
includes a software recovery procedure for Lock Release.
Follow this procedure to start the motor controller test:
1. Power up the Ultrasound system.
2. After initialization is complete, press CONFIG... on the
Utility tab on the Touch Panel.
3. Log on as ADM.
4. Select System from the bottom of the menu that appears on
the monitor.
5. Select Test.
Figure 7-2. Test screen
Vivid E80/E90/E95
Service Manual
GC091052 Rev. 3