System Setup
To install a Software Option (continued)
CAUTION
3-38
4. Select New to open the New Key dialog where you type the
SW Option Key.
Incorrect Software Option Key entry will result in loss of
functionality. If Software Option Key is not accepted by the
Ultrasound system, please contact your local GE Service
Representative or the Online Center.
5. Type the Software Option Key. It is case sensitive, and you
must include the dashes (-) as they are part of the Software
Option Key.
6. Press Save to save the new setting.
7. Restart to save and activate the settings and adjustments
you have done so far.
Vivid E80/E90/E95
GC091052 Rev. 3
Service Manual