Profiling tab

Burst sampling
Figure 2-12 Profiling tab
10. Profiling Tab (Figure 2-12)
Select the Data Interval (recording interval) in minutes
from the dropdown list.
11. Select the Samples per Interval (measurement interval).
From the drop-down list, select the desired number of mea-
surements per Data Interval. A larger number will reduce
battery life but improve measurement precision. Teledyne
Isco recommends using no more than 5 Samples per Inter-
val in most applications.
12. Burst Sampling
Burst sampling (reading interval) is a secondary data
interval automatically triggered by a programmed thresh-
old condition, such as a rain event or overflow.
13. To enable Burst Sampling, check the box next to Burst
Sampling (Figure 2-12), then select Configure.
a. In the Configure window that appears, select the data
interval in minutes, and the number of samples (read-
ings) to be taken when the specified trigger conditions
are met.
ADFM® Flow Meter
Section 2 Setup and Installation
2-13