Forced Depth And Scaling Constant (Hot Tap Only) - Teledyne ADFM Pro20 Installation And Operation Manual

Acoustic doppler flow meters
Hide thumbs Also See for ADFM Pro20:
Table of Contents

Advertisement

ADFM® Flow Meter
Section 2 Setup and Installation
2.7.1 Forced Depth and
Scaling Constant (Hot
Tap only)
Early version HTI
Current HTI
Figure 2-43 Location of the QS value on Hot Tap label
2-40
During Hot Ta p site setup (F lowlink) or statio n setup
(WinADFM), you must open the Advanced Commands field and
enter the following information:
Forced Depth (QF) – The value of QF is the pipe's inside
diameter (ID) in millimeters.
Example: For an ID of 22.5 inches, enter QF572 (22.5 x 25.4
mm/in, rounded up).
Scaling Constant (QS) – This is the number printed on the
serial tag of every Hot Tap sensor. Enter as QS### or QS####,
depending on how many numbers are given.
Example: Figure 2-43 shows the location of this value on old
(QS1005) and new (QS982)sensors.