ADFM® Flow Meter
Section 2 Setup and Installation
Channel tab
Figure 2-9 Channel Tab window
2-10
5. Channel Tab
Select the application geometry and enter the appropriate
dimensions. For detailed information about determining
channel geometry, see Section 2.6.1 Pro20: Channel Geom-
etry.
a. Bed Level is the level of permanent silt/debris in the
bottom of the channel.
b. Zero Offset is the distance from the bottom of the
ADFM sensor to the bottom of the channel. If you are
using the Flow Conditioning Platform (FCP), this value
is typically 0.24 inches (6.1 mm).
c. FCP Height is the height or thickness of the Flow Con-
ditioning Platform. It is typically 0.88 inches (22.5
mm).
If the channel type is not irregular, skip to step 9.