Setting The Automatic Defrost Time (Cryochamber); Programming The Temperature Of The Cryochamber - Leica CM1520 Instructions For Use Manual

Cryostat
Hide thumbs Also See for CM1520:
Table of Contents

Advertisement

6
Operating the Instrument

6.2.2 Setting the automatic defrost time (cryochamber)

Fig. 16
The automatic defrost cycle takes place once every 24 hours.
Briefly touch the Plus
defrost time that is currently set. The two LEDs between the display of hours and minutes flash at the
same time.
• To change the beginning of the defrost time in 15-minute increments, touch or press and hold the
or
Minus
button. The defrost duration is 12 minutes.

6.2.3 Programming the temperature of the cryochamber

2
Fig. 17
The temperature of the cryochamber is set and indicated on the panel marked with the cryostat symbol.
The actual temperature is the standard indication. Briefly touch the Plus
(→ "Fig. 17-2")
• You can now set the desired value using these buttons. Pressing and holding the
continuously increases or decreases the cryochamber temperature.
• 5 seconds after finishing the programming the current value will be displayed again.
32
2
1
(→ "Fig. 16-1")
or
1
button to display the target temperature.
Minus
(→ "Fig. 16-2")
button to display the beginning of the
(→ "Fig. 17-1")
or
Minus
Plus
or
Minus
button
Version 2.0, Revision K
Plus