Leica CM1520 Instructions For Use Manual page 51

Cryostat
Hide thumbs Also See for CM1520:
Table of Contents

Advertisement

The anti-roll guide system can be moved sideways (only for the 84 mm variant). To better find the mid
position, a groove
sits.
• The CN knife holder can be used to hold tungsten-carbide or steel knives.
19
Fig. 46
Warning
The safety guard must never be uninstalled!
Clearance angle adjustment
The clearance angle adjustment for the knife holder CN is carried out exactly as described for the blade
holder CE (→ P. 40 – 7.4.2 Clearance angle adjustment).
Adjusting the anti-roll guide
The anti-roll guide adjustment for the knife holder CN is carried out exactly as described for the blade
holder CE (→ P. 46 – Adjusting the anti-roll guide system).
To move it laterally, see (→ "Fig. 46").
Knife holder CN with anti-roll guide – Moving the clamping jaws
Note
If large specimen discs are to be used with the knife holder (e.g. 50 x 80 mm), it is possible to shift the
clamping jaws laterally.
The clamping jaws are factory-installed in the knife holder with a clearance of 64 mm. If necessary,
both clamping jaws can be set with a clearance of 84 mm.
Leica CM1520
(→ "Fig. 46-17")
is provided in the shaft
18
17
Daily Use of the Instrument
(→ "Fig. 46-16")
19
16
in which the anti-roll guide
7
51