Leica CM1520 Instructions For Use Manual page 23

Cryostat
Hide thumbs Also See for CM1520:
Table of Contents

Advertisement

• Carefully roll the instrument backwards over the ramp from the pallet
castors to the installation location.
Transport to the site
• Check if the location meets the requirements specified in the
(→ P. 20 – 4.1 Installation site requirements).
• Transport the instrument to the desired location and observe the following notes:
Warning
• The instrument must be transported in either an upright position or slightly tilted (max. 30 °).
• When tilting the instrument 2 people must counterbalance from the front side as the instrument can
tip over and cause severe injury, as well as severe damage to the instrument and the surrounding
area!
• When transporting the instrument on the castors (→ "Fig. 8-2"), grip the instrument only at the marked
housing locations (→ "Fig. 8-3").
• To ensure a secure upright position at its intended location, the instrument must be adjusted.
To do so, unscrew both adjusting feet (→ "Fig. 8-1")
(→ "Fig. 8-4").
• When subsequently transporting the instrument any further on castors, screw them back in all the way.
4
Fig. 8
Leica CM1520
2
1
Installing the Instrument
using the provided open-end wrench (size 13)
3
2
1
(→ "Fig. 7-2")
and slide it on the
3
2
4
23