Daily Use of the Instrument
Adjusting the anti-roll guide system
You can adjust the height of the anti-roll guide system using the knurled nut
• If you turn the nut counterclockwise, the anti-roll guide system moves toward the blade.
• If you turn the nut clockwise, the anti-roll guide system moves away from the blade.
If the anti-roll guide system is in the wrong position relative to the blade, the following problems will
result:
» The section rolls over the glass insert of the anti-roll guide system (→ "Fig. 39-1").
L Error: Glass insert not high enough.
 Remedy: Turn the knurled nut counterclockwise until the section is pushed between the blade and
anti-roll guide as shown
Fig. 38
» Section are compressed and block hits the glass insert
L Error: Anti-roll guide system is set too high.
 Remedy: Turn the knurled nut clockwise until the section is pushed between the blade and anti-roll
guide as shown in (→ "Fig. 39-3").
1
Fig. 39
46
in
(→ "Fig. 39-3").
8
2
(→ "Fig. 38-8"):
(→ "Fig. 39-2")
after sectioning.