11 Notes on EMC

 Medical electrical equipment is subject to special precautions with regard to EMC
and must be installed according to the following EMC notes.
Guidance and manufacturer's declaration – ambient conditions
The ATMOS C 11 Systema is suitable for use in the following environments:
• In professional healthcare facilities such as medical practices, hospitals/clinics, first-
aid facilities, and operating theatres/rooms.
The following environments are not suitable:
Vicinity of HF surgical devices and in settings outside of an HF-shielded room of a mag-
netic resonance imaging system.
The customer or user of the ATMOS C 11 Systema must ensure that the device is used in
a prescribed environment.
Guidance and manufacturer's declaration – key features
 Please note the technical data in these instructions. The essential features are fully
usable even in the presence of electromagnetic disturbances.
Guidance and manufacturer's declaration – removable components that can be
replaced by the operator
The ATMOS C 11 Systema has the following removable components that can be replaced
by the operator:
Type
ATMOS LS 21 LED
ATMOS LS 21 LED, warm white
ATMOS HL 21 LED
Power cable
Guidance and manufacturer's declaration – warnings
WARNING
The use of electrical components and accessories other than those specified or provid-
ed by the manufacturer may cause increased electromagnetic interference or reduced
immunity to electromagnetic interference and result in faulty operation of the device.
WARNING
Portable RF communications equipment (e.g., radios, antenna cables) should be used no
closer than 30 cm* to any part of the ATMOS C 11 Systema, including cables, specified
by the manufacturer. Otherwise, degradation of the performance of this equipment
could result.
 *The distance may be reduced at higher immunity test levels.
REF
Max. cable length
541.4300.0
2.0 m
541.4400.0
2.0 m
530.4020.0
2.0 m
507.0859.0
3.0 m