13 Notes on EMC

 Medical electrical equipment is subject to special precautions with regard to EMC
and must be installed according to the following EMC notes.
Guidance and manufacturer's declaration – ambient conditions
The ATMOS C 451 is suitable for use in the following environments:
• In professional healthcare facilities such as medical practices, hospitals/clinics, first-
aid facilities, and operating theatres/rooms.
It is not suitable for use in the vicinity of HF surgical devices and in settings outside
of an HF-shielded room of a magnetic resonance imaging system.
• Special environments such as factory or military facilities and medical areas near HF
surgical devices, short-wave therapy equipment, or within an HF-shielded room of a
magnetic resonance imaging system.
The customer or user of the ATMOS C 451 must ensure that it is used in such a pre-
scribed environment.
Guidance and manufacturer's declaration – key features
 Please note the technical data in this manual. The essential features are fully usable
even in the presence of electromagnetic disturbances.
Guidance and manufacturer's declaration – electrical components
The ATMOS C 451 has the following electrical components:
Type
Power cable
Guidance and manufacturer's declaration – warnings
Warning
The use of electrical components and accessories other than those specified or provid-
ed by the manufacturer may cause increased electromagnetic interference or reduced
immunity to electromagnetic interference and result in faulty operation of the device.
Warning
Portable HF communications equipment (e.g., radios, antenna cables) should not be
used within 30 cm* of any parts as specified by the manufacturer or of the ATMOS C 451
cables. Failure to do so may lead to a reduction in the device's performance.
 *The distance may be reduced at higher immunity test levels.
Warning
Placement on top of or next to another device should be avoided. This could result in
faulty operation. If such placement cannot be avoided, the proper functioning of the
device must be monitored regularly. If possible, please switch off any nearby devices
that are not in use.
REF
Max. cable length
008.0818.0
5 m
Notes on EMC