9.0 Technical data

Voltage
Special voltage
Current consumption
Power consumption
Connections
Fuses
Stimulation time
Timer indication
Water temperature
Temperature indication
Feed temperature of the water
Water quality
Water flow in the irrigation mode
(high flow)
Water flow in the stimulation mode
(low flow)
Operating time
Operating pressure
Protective earth conductor resistance
Earth leakage current
Enclosure leakage current
Patient leakage current
Ambient conditions
Transport / storage
Ambient conditions
Operation
Maximum operational altitude
Contamination level
Overvoltage category
Dimensions HxWxD
Weight
Period tests
Safety class (EN 60601-1)
Degree of protection
Protection class
Further classifications according to other regu-
lations
Classification according to Appendix IX EC
Directive 93/42/EEC
CE marking
20
220 - 240 V~ ± 10 %; 50/60 Hz
100 - 127 V~ ± 10 %; 50/60 Hz
max. 6.8 A (220 - 240 V~); 14.8 A (100- 127 V~)
max. 1500 W
Power connection via IEC socket; control output for a nystagmograph; equi-
potential equalization; water supply G 3/4" external thread; water drainage
G 1/4" external thread; connection for double-barrelled hose
F 8 A (f. 220 - 240 V~); F 15 A (f. 110 - 127 V~)
Adjustable by timer from 1 up to 99 sec.
Indication accuracy ± 0.5 s ± ½ digit
20°C - 47°C
Indication accuracy ± 0.6°C ± ½ digit (can only be guarantee with un-
changed ambient conditions)
+7...+19.5 °C
Drinking water
450 ml/min: Standard unit and variant 250/450 ml/min
400 ml/min: Variant 400/400 ml/min
500 ml/min: Variant 500/500 ml/min
150 ml/min: Standard unit 150/450 ml/min (adjustable)
250 ml/min: Variant 250/400 ml/min (adjustable)
400 ml/min: Variant 400/400 ml/min (adjustable)
500 ml/min: Variant 500/500 ml/min (adjustable)
Continuous operation
At least: +2000 hPa (2 bar); max.: +6000 hPa (6 bar)
max. 0,1 Ω
max. 0.5 mA
max. 0.1 mA
max. 0.1 mA
-20...+50°C
5...90 % air humidity without condensation
air pressure of 700...1060 hPa
+10...+35°C
20...80 % air humidity without condensation
air pressure of 700...1060 hPa
≤ 3000 m (NN)
Class 2
II
14.5 x 37 x 31.5 cm
Approx. 5,5 kg
Inspection according to the manufacturers specifications every 12 months.
I
Type B
IPX0
VDE protection class 1 (IEC 601/EN 60601)
Class IIa
CE 0124