11.0 Technical data
11.1 ATMOS S 201 Thorax
Input voltage
Power consumption
Rechargeable battery, built-in
Fuses
Other safety lugs
Pump performance
Display
Data memory
Mode of operation
Battery operation time at
maximum continuous suction
Battery operation time at
standard operation (without
fistula)
Battery recharging time
Earth leakage current
Patient leakage current
Environmental conditions
Transport/storage
• Temperature range
• Air humidity without
condensation
• Air pressure
Environmental conditions
Operation
• Temperature range
• Air humidity without
condensation
• Air pressure
Maximum operating altitude
Contamination level
Overvoltage category
Dimensions (H x W x D)
Weight
Housing material
100 – 240 V~; 50/60 Hz
max. 70 VA
Li-Ion, 14.4 V nominal, 3350 mAh nominal
1 x T 1.25 A/H, 250 V
Pressure control valve „pop-off" in the canister.
Vacuum limitation in the device to approx. 150 mbar.
Acoustic and optical error warnings.
Freeflow 18 ± 2 l/ min.
Vacuum adjustable from −5 mbar to −100 mbar, step size
−1 mbar.
Graphic display, colour, with background lighting, display of
target vacuum and actual vacuum in mbar, cmH
flow in ml/min or l/min
Internal memory for therapy data: 2.5 MB.
Up to 12 days recording possible.
Continuous operating, at ambient temperatures.
Simultaneous battery recharging and operation possible.
1 h
12 h
Fully recharged (at least 95 %) in approx. 2.5 h
max. 0.5 mA
max. 0.01 mA
−10...+50 °C
30...95 %
700...1060 hPa
+10...+35 °C
30...95 %
700...1060 hPa
3000 m (NN)
Class 2
II
approx. 365 x 250 x 168 mm
3.7 kg (device with canister)
ABS/PC UL 94 V0, grey-white and pigeon blue
O, kPa and
2
61