EDAN SE-601 Series User Manual page 56

Electrocardiograph
Hide thumbs Also See for SE-601 Series:
Table of Contents

Advertisement

The above figures (a) and (b) show the ECG reports in the AUTO mode. Template is set to
On, and Auto Style is set to 6×2+1rhy.
Figure (a) shows:
ID: 0000004322 (Patient ID)
10-10-2008 16:00:43 (Current Date & Current Time)
Gender: Male, Age: 25 years, Height: 167cm, Weight: 64kg, BP: 112/78mmHg
(1mV calibration mark)
І, II, Ш, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6 (12 standard Leads) and ECG waves
II on the bottom (Rhythm Lead) and rhythm wave
0.15~100Hz (0.15Hz DFT Filter, 100Hz Lowpass Filter)
AC50 (50Hz AC Filter)
25mm/s (Paper Speed)
10mm/mV (Gain)
60 (Heart Rate)
V1.0 (Version number)
Note: The version number in the ECG reports is just for reference. The real version
number can be found on the screen of the electrocardiograph when you turn
on the device.
Figure (b) shows: Patient Information, Measure Information, Minnesota Code, Diagnosis
Information, Average Template, Report Reviewed by, Doctor Name, Current Date and Current
Time, 0.15Hz (DFT Filter), 100Hz (Lowpass Filter), AC50 (50Hz AC Filter), 25mm/s (Paper
Speed), 10mm/mV (Gain), V1.0 (Version number).
Patient Information includes:
ID, Name, Age, Gender, Height, Weight, BP, Race, Medication, Ward NO, Tel (Extra
Information)
Measure Information includes:
HR (Heart Rate)
P Dur----P wave duration: the average P-wave duration from several selected dominant beats;
PR int----P-R interval: the average P-R interval from several selected dominant beats;
QRS Dur----QRS complex duration: the average QRS complex duration from several selected
dominant beats;
QT/QTC int----Q-T interval: the average Q-T interval from several selected dominant beats
Normalized QT interval;
P/QRS/T axis----Dominant direction of the average integrated ECG vectors;
RV5/SV1 amp----The maximum value of the amplitude of R and R' wave from the average beat
of lead V5 / The maximum absolute value of the amplitude of S and S' wave
from the average beat of lead V1;
RV5+SV1 amp----Sum of RV5 and SV1;
RV6/SV2 amp----The maximum value of the amplitude of R and R' wave from the average beat
SE-601 Series Electrocardiograph User Manual