EDAN SE-601 Series User Manual page 106

Electrocardiograph
Hide thumbs Also See for SE-601 Series:
Table of Contents

Advertisement

SE-601 Series Electrocardiograph User Manual
Q4: I want to print the hospital name in the report, but I can't find the place to enter it, where is
it?
A4: Please open the More Setup interface, and move the cursor to the Institution textbox, and
then input the hospital name. The content you inputted in this textbox will be printed in the report.
For details, please refer to Section 9.9.4, "Entering Institution".
Q5: I pressed the START/STOP key, but the ECG didn't start printing, what's wrong with it?
A5: Please check whether there is any error information displayed on the screen.
If the hint No Paper or Paper Error is shown on the screen, please deal with it according to the
above-mentioned measures.
If the hint Transmitting... is shown on the screen, which means that the ECG is transmitting the
data to the PC, please wait a few seconds. You can start the printing after the data has been
transmitted.
If none of the above-mentioned information is shown on the screen, please check whether
Sample Mode is set to Pre-Sample on the Work Mode Setup interface. If it is so, the system
will not respond to the START/STOP key unless 10s data has been collected. In this case, all you
have to do is to wait a few seconds, and then you are able to start the printing by pressing the
START/STOP key again.
If the problem still exists, please contact the manufacturer or the local distributor for further
disposal.
3. Transmitting Problems
Q1: The ECG doesn't respond to any keys after a long time of transmitting. It transmits nothing
for there is no new data appearing on the interface of the PC software. What should I do?
A1: Some error may occur during the transmission course, for example, the connection between
the ECG and the net cable may loosen. In this case, please restart the ECG. If it doesn't work,
please restart the PC.
If the problem still exists, please contact the manufacturer or the local distributor for further
disposal.
4. Main Unit Problems
Q1: After power-on, the ECG stays on the logo interface and doesn't open the main interface. I
have reset the machine several times, but there is no better change.
A1: The reason for this problem might be: there is a key pressed down, without springing up.
Find that key, and make it spring up, the problem should be solved.
- 99 -