Disposable Electrode Connection
8)
Align all lead wires of the patient cable to avoid twisting, and connect the alligator clips
to the lead wires.
9)
Clean the electrode areas on the body surface with 75% alcohol.
10) Attach the disposable electrodes to the electrode positions on the body surface.
11) Clip the disposable electrodes with the alligator clips.
The quality of ECG waveform will be affected by the contacting resistance between the patient
and the electrode. In order to get a high-quality ECG, the skin-electrode resistance must be
minimized while connecting electrodes.
:
WARNING
1. Make sure that all electrodes are connected to the patient correctly before
operation.
2. Ensure that the conductive parts of electrodes and associated connectors,
including neutral electrodes, do not come in contact with earth or any other
conducting objects.
: The disposable electrodes can only be used for one time.
CAUTION
SE-601 Series Electrocardiograph User Manual
V1
V2
V3
- 35 -
V4
V6
V5