EDAN SE-601 Series User Manual page 40

Electrocardiograph
Hide thumbs Also See for SE-601 Series:
Table of Contents

Advertisement

C3: Fifth rib between C2 and C4
C4: Fifth intercostal space on the left midclavicular line
C5: Left anterior axillary line at the horizontal level of C4
C6: Left midaxillary line at the horizontal level of C4
C1
C2
C3
Chest Electrode Connection:
1)
Ensure that the electrodes are clean;
2)
Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to
the corresponding electrodes according to the colors and identifiers;
3)
Clean the electrode area on the chest surface with 75% alcohol;
4)
Daub the round area of 25mm in diameter on each electrode site with gel evenly;
5)
Place a small amount of gel on the brim of chest electrode's metal cup;
6)
Place the electrode on the chest electrode site and squeeze the suction bulb. Unclench it
and the electrode is adsorbed on the chest;
7)
Attach all chest electrodes in the same way.
Limb Electrode Connection:
1)
Ensure that the electrodes are clean;
2)
Align all lead wires of the patient cable to avoid twisting, and connect the lead wires to
the corresponding electrodes according to the colors and identifiers;
3)
Clean the electrode area which is a short distance above the ankle or the wrist with 75%
alcohol;
4)
Daub the electrode area on the limb with gel evenly;
5)
Place a small amount of gel on the metal part of the limb electrode clamp;
6)
Connect the electrode to the limb, and make sure that the metal part is placed on the
electrode area above the ankle or the wrist;
7)
Attach all limb electrodes in the same way.
SE-601 Series Electrocardiograph User Manual
- 33 -
C4
C6
C5