Lower Unit Repairs (Bedded and Freestanding Models)
6.
Install new foot switch. Make sure that you do not pinch the electrical wires.
7.
Ensure that foot switch connectors are secured with retaining clips.
5.7.4 Caster Replacement
The casters may be replaced with the unit upright or carefully placed on its back.
1.
Lock all the other casters.
2.
Lift the caster off the floor and use blocks to support the leg near the caster you are replacing.
3.
Remove screw(s) from boss underneath the leg casting.
4.
Secure the caster to the leg by tightening the screw(s) through the caster plate into the boss in the
bottom of the leg.