Bench Alignment With Alignment Tool - Otto Bock 7E10 Helix 3D Instructions For Use Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 24

3.4 Bench alignment with alignment tool

 CAUTION
Danger as a result of incorrect alignment. Errors during the alignment of the prosthesis can lead to malfunc­
tion of the joint up to loss of function due to structural failure. This can cause the patient to fall.
During bench alignment, it is essential to make sure that the hip joint is fully extended!
Preparing the hip joint for bench alignment:
To facilitate alignment in the L.A.S.A.R. Assembly, please remove the integrated expansion springs as described
in Section 3.8.
Fix the hip joint in full extension using the enclosed cable tie. Position the cable tie approximately at the level where
the lower section ends (Fig. 2).
Helix
Position of
cable tie
1
2
Ottobock | 35

Advertisement

Table of Contents
loading

Table of Contents