10 Instrument Functions
A
A
ENDO MAX. 4%
24
%
%
Sirona Dental Systems GmbH
10.3

SIROSON L

Once the SIROSON L handpiece has been
removed and started, the programmed
intensity value in % and the operator A or B
appear on the display.
The intensity is set with the
1% to 100%.
If you hold the corresponding key down,
the intensity is adjusted in counting steps:
1%: 1 – 4%
5%: 10% – 100%
Fast adjustment:
If you briefly touch the relevant key, the
speed preselection always jumps to the next
step of 20% up or down.
The cooling water flow is always switched
on automatically.
SIROSON L − endodonty
10.4
The intensity range of 1-4% is provided for
endodontics. It is adjustable in steps of 1%.
The warning ENDO MAX. 4% appears on
the display in this case
For reasons of security, we recommend
that you program the set value for endo-
donty. Otherwise, the value programmed
until now becomes active again during
storing the instrument in the carrier in the
meantime.
ATTENTION
If 4% is exceeded during the
endodonty treatment, temperature dam-
ages and needle breaks can arise.