COOLING AIR
41 63 189 D 3311
D 3311.201.01.10.02
A
TAKE INSTRUMENT
bar
A
NO COOLING AIR
A
# FUNCTION
TOGGLE - MOMENTARY
A
PURGE TIME
120 SEC
12 Basic Workstation Settings with Mode Key
12.11

Mode: COOLING AIR

Cooling air intensity for SIRONA motor
First, display of the request for removing a
SIRONA motor instrument.
After pressing the + or – button, the preset
cooling air pressure is displayed; value:
2 bar - 3.5 bar. 3.4 bar ± 0.1 bar must be set
for the SIRONA motor. Correct, if
necessary, by pressing the +/ – buttons.
ATTENTION: If the cooling air is not
sufficient, the motor is automatically
switched off. The message NO
COOLING AIR will flush on the display.
Please call service technicean.
After setting down the motor, please adjust
the next motor. You can leaf forwards in the
Mode menu only after setting the cooling air
intensity for all motors.
Since no cooling air setting is required for
the turbine or the SPRAYVIT,
NO COOLING AIR appears on the display
when these instruments are removed.
12.12
Mode: #FUNCTION
TOGGLE - MOMENTARY
Call key # as push button or switch
(connected by technician).
In this mode, you can define call key # as a
pushbutton (MOMENTARY) or as a switch
(TOGGLE)
(possible only when the SIVISION Mode
preselection is set to NO).
12.13

Mode: PURGE TIME..SEC

Setting the purge time for
flushing the water paths
(Only for dentist unit)
The purge time per instrument can be
preset for the Purge function
(see page 31) from 60 to 120 seconds with
– +
the
keys.
35