Performance tests
46
8. In the Channel tab under Setup, select Channel 1 and set the Output Path to
DC High BW.
9. Set the amplitude of the instrument as shown in the table for the Output Path
under test. (See Table 34.)
Table 34: Analog amplitude accuracy (DC High BW output path)
Amplitude settings
DC High BW output path
25 mV
p-p
100 mV
p-p
200 mV
p-p
500 mV
p-p
1 V
p-p
1.5 V
p-p
(Option DC)
DC High Volt output path
(Option HV)
10 mV
p-p
100 mV
p-p
500 mV
p-p
5 V
10. In the Setup tab, enable the channel's output.
11. Click the Play button on-screen or press the button on the front panel of the
generator.
12. Press the All Outputs Off button on the generator (toggle from red to green)
to enable all outputs.
Accuracy limits
23.75 mV to 26.25 mV
98 mV to 102 mV
196 mV to 204 mV
480 mV to 520 mV
980 mV to 1.02 V
1.47 V to 1.53 V
9.5 mV to 10.5 mV
98 mV to 102 mV
480 mV to 520 mV
4.92 V to 5.08 V
AWG5200 Series Technical Reference