Tektronix AWG5200 Series Technical Reference page 59

Arbitrary waveform generators, specifications and performance verification
Hide thumbs Also See for AWG5200 Series:
Table of Contents

Advertisement

AWG5200 Series Technical Reference
13. Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_1.
14. Use the following formula to compensate the voltage for the 50 Ω BNC
termination:
V_high = [(Term_R + 50) / (2 Term_R)] Measured_voltage_1
Where Term_R is the resistance of the 50 Ω BNC termination. (See page 44,
Termination resistance measurement.) procedure.
15. From the Waveform List window, assign the waveform PV_DC_Minus.wfmx
to Channel 1.
16. Measure the output voltage on the digital multimeter and note the value as
Measured_voltage_2.
17. Use the following formula to compensate the voltage for the 50 Ω BNC
termination:
V_low = [(Term_R + 50) / (2 Term_R)] Measured_voltage_2
Where Term_R is the resistance of the 50 Ω BNC termination. (See page 44,
Termination resistance measurement.) procedure.
18. Verify that the voltage difference |(V_high-V_low)| falls within the limits
given in the table. (See Table 34 on page 46.)
19. Repeat steps 9 through 18 for each amplitude setting in the table. (See
Table 34 on page 46.)
20. Move the SMA-BNC adapter from the CH 1 (+) connector to the CH 1 (–)
connector and move the 50 Ω SMA termination from the CH 1 (–) connector
to the CH 1 (+) connector.
21. Repeat steps 9 through 19 for the CH1 (–) connector.
22. Repeat steps 7 through 21 until all channels are checked, modifying the
instructions with channel number under test.
Performance tests
47