Analog offset accuracy (DC output paths)

AWG5200 Series Technical Reference
28. Repeat steps 9 through 19.
29. Repeat steps 7 through 28 until all channels are checked, modifying the
instructions with channel number under test.
30. Press the All Outputs Off button on the generator (toggle from green to red)
to disable all outputs.
31. Disconnect the test setup.
Required equipment
Digital multimeter
BNC-dual banana adapter
50 Ω BNC termination
SMA female-BNC male adapter
50 Ω SMA termination
Before starting this procedure, ensure you have the "Term R" value used in the
calculations. (See page 44, Termination resistance measurement.)
1. Press the All Outputs Off button on the generator (toggle from green to red)
to disable all outputs.
2. Connect an SMA-BNC adapter to the 50 Ω BNC termination on the digital
multimeter.
3. Connect the CH 1 (+) connector from the generator to the HI and LO inputs of
the digital multimeter using a 50 Ω SMA cable.
4. Terminate the CH 1 (–) connector on the generator using a 50 Ω SMA
terminator.
Performance tests
Prerequisites
Instrument preparation and load test
waveforms(See page 42, Prerequisites.)
Termination resistance measurement
procedure
49