Tektronix AWG5200 Series Technical Reference page 53

Arbitrary waveform generators, specifications and performance verification
Hide thumbs Also See for AWG5200 Series:
Table of Contents

Advertisement

AWG5200 Series Technical Reference
7. Click the Home tab on the display.
8. Click the Reset to default setup button in the toolbar.
9. Set the function generator to output a 1 kHz square wave at 5 V
10. Turn on the output of the function generator.
11. From the Waveform List window, assign the waveform PV_Square.wfmx to
the channel under test.
12. On the Home tab, set the generator's channel 1 Run Mode to Triggered and
set the Trigger Input to A.
13. In the Setup tab, click Channel, and enable the channel's output.
14. In the Setup tab, click Trigger, and set the trigger level to 1.0 V, Rising, 50 Ω.
15. Click the Play button on-screen or on the front panel of the generator.
16. Press the All Outputs Off button on the generator (toggle from red to green)
to enable all outputs.
17. Click the Home tab and verify that the output is displayed on the generator.
18. Verify that the output is displayed on the test oscilloscope.
19. Repeat the test for the Trigger B input after making the following changes.
Move the cable from the Trigger A input to the Trigger B input.
On the Home tab, set the trigger input to B.
20. Disconnect the test setup.
Brief procedures
.
p-p
41