Tektronix AWG5200 Series Technical Reference page 60

Arbitrary waveform generators, specifications and performance verification
Hide thumbs Also See for AWG5200 Series:
Table of Contents

Advertisement

Performance tests
48
23. Press the All Outputs Off button on the generator (toggle from green to red)
to disable all outputs.
NOTE.
This is the start of testing the optional DC High Voltage output path.
If the optional DC High Voltage output is licensed, continue with this procedure.
If not, skip to step 31.
24. In the Channel tab under Setup, select Channel 1 and set the Output Path to
DC High Volt.
25. Move the SMA-BNC adapter from the CH 1 (–) connector to the CH 1 (+)
connector and move the 50 Ω SMA termination from the CH 1 (+) connector
to the CH 1 (–) connector.
26. Repeat steps 9 through 18 for each amplitude setting in the table for the DC
High Volt output path. (See Table 34 on page 46.)
27. Move the SMA-BNC adapter from the CH 1 (+) connector to the CH 1 (–)
connector and move the 50 Ω SMA termination from the CH 1 (–) connector
to the CH 1 (+) connector.
AWG5200 Series Technical Reference

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Awg5204Awg5202Awg5208

Table of Contents