Warning
The servicing instructions are for use by qualified
personnel only. To avoid personal injury, do not
perform any servicing unless you are qualified to
do so. Refer to all safety summaries prior to
performing service.
www.tektronix.com

Advertisement

Table of Contents
loading

Summary of Contents for Tektronix AWG5000 Series

  • Page 1 Service Manual AWG5000 Series Arbitrary Waveform Generators 071-2083-00 Revision A Warning The servicing instructions are for use by qualified personnel only. To avoid personal injury, do not perform any servicing unless you are qualified to do so. Refer to all safety summaries prior to performing service.
  • Page 2 Copyright © Tektronix. All rights reserved. Licensed software products are owned by Tektronix or its suppliers and are protected by United States copyright laws and international treaty provisions. Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes that in all previously published material.
  • Page 3 Tektronix, with shipping charges prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the Tektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations.
  • Page 5: Table Of Contents

    Returning the Instrument for Service........4-56 AWG5000 Series Service Manual...
  • Page 6 Using the Replaceable Parts List......... . 5-3 AWG5000 Series Service Manual...
  • Page 7 List of Figures Figure 2-1: AWG5000 Series block diagram ..... . . 2-2 Figure 3-1: Service UI menu ........3-3 Figure 3-2: 10 MHz Reference clock calibration test hookup .
  • Page 8 Figure 5-14: Exploded diagram - BNC bracket ....5-25 Figure 5-15: Exploded diagram - Rear connectors ....5-26 AWG5000 Series Service Manual...
  • Page 9 Table 4-10: Hardware error ........4-53 AWG5000 Series Service Manual...
  • Page 10 List of Tables AWG5000 Series Service Manual...
  • Page 11: General Safety Summary

    Do Not Operate With Suspected Failures. If you suspect there is damage to this product, have it inspected by qualified service personnel. Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present. Do Not Operate in Wet/Damp Conditions. Do Not Operate in an Explosive Atmosphere. AWG5000 Series Service Manual...
  • Page 12 CAUTION indicates a hazard to property including the product. Symbols on the Product. The following symbols may appear on the product: WARNING Protective Ground CAUTION Double High Voltage (Earth) Terminal Refer to Manual Insulated viii AWG5000 Series Service Manual...
  • Page 13 Use Care When Servicing With Power On. Dangerous voltages or currents may exist in this product. Disconnect power, remove battery (if applicable), and disconnect test leads before removing protective panels, soldering, or replacing components. To avoid electric shock, do not touch exposed connections. AWG5000 Series Service Manual...
  • Page 14 Service Safety Summary AWG5000 Series Service Manual...
  • Page 15: Environmental Considerations

    European Union's requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com). Mercury Notification. This product uses an LCD backlight lamp that contains mercury.
  • Page 16 Environmental Considerations AWG5000 Series Service Manual...
  • Page 17: Modules (Top)

    I/O port connectors, that is listed in the replaceable parts list of this manual. Safety Symbols and terms related to safety appear in the Service Safety Summary found at the beginning of this manual. AWG5000 Series Service Manual xiii...
  • Page 18 Preface AWG5000 Series Service Manual...
  • Page 19 Operating Information...
  • Page 21: Operating Information

    Operating Information For information on installing, operating, and networking the instrument, refer to the AWG5000 Series Arbitrary Waveform Generators Quick Start User Manual. This manual is available on the Document CD that came with your instrument, and on the Tektronix Web site (www.tektronix.com/manuals).
  • Page 22: Awg5000 Series Service Manual

    AWG50UP Option M14: AWG5014 Field upgrade for waveform length expansion from 16 M to 32 M points AWG50UP Option D03: AWG5002 Field upgrade for digital data outputs AWG50UP Option D13: AWG5012 Field upgrade for digital data outputs AWG5000 Series Service Manual...
  • Page 23 Theory of Operation...
  • Page 25: Theory Of Operation

    Theory of Operation This section describes the electrical operation of the AWG5000 Series Arbitrary Waveform Generators. Overview The AWG5000 Series Arbitrary Waveform Generators provide four models with different frequency and numbers of channels. Each model consists of two major sections: the platform section and generator section. The platform section is common to each model.
  • Page 26: Figure 2-1: Awg5000 Series Block Diagram

    Theory of Operation Figure 2-1: AWG5000 Series block diagram AWG5000 Series Service Manual...
  • Page 27: Module Overviews

    An EEROM on this board is used to store the instrument setting and calibration data. The MIO board consists of the following blocks: PCI controller GPIB controller LCD display controller (ATI Mobility-MI) PCI target adapter PLD140 (Xilinx FPGA) EEPROM for calibration data storage TekLink interface AWG5000 Series Service Manual...
  • Page 28 Three fans on the side of the instrument provide forced air cooling. The fans are controlled by the fan speed control circuit on the PWR board and are regulated by monitoring the temperature at the circuitry. AWG5000 Series Service Manual...
  • Page 29 PLD120 (Xilinx Virtex-4 FPGA) as a memory controller PLD121 (Xilinx Virtex-4 FPGA) as an AWG controller ZBT type SRAM for the waveform memory 28 bits digital output driver Marker output driver Trigger and event inputs DC output DC-DC converter AWG5000 Series Service Manual...
  • Page 30 Amplitude, filter, offset, add input, and On/Off controls are added to the analog outputs. The OUT1G board consists of the following blocks: Analog differential output amplifier Selectable low-pass filter Selectable attenuator Add input from external Control D/A for offset setting Level monitor A/D Relay driver Controller PLD AWG5000 Series Service Manual...
  • Page 31 Adjustment Procedures...
  • Page 33: Adjustment Procedures

    30 °C. Adjustments performed before warm-up or outside this temperature range may result in poor performance. Adjustment Interval If the instrument fails performance tests (refer to the AWG5000 Series Arbitrary Waveform Generators Technical Reference Manual), then adjustment may be required.
  • Page 34: Table 3-1: Test Equipment

    You must enable the Service menu to perform the adjustment procedure. To do this: 1. Power on the instrument. 2. Select the System menu from the menu bar, and then select Service Mode... 3. The following dialog box is displayed. Enter the password “1185”. AWG5000 Series Service Manual...
  • Page 35: Figure 3-1: Service Ui Menu

    You must finish a calibration item listed on the upper part of the screen before performing the lower part of the items. Some items may be grayed out, depending on your instrument option configurations. AWG5000 Series Service Manual...
  • Page 36 A warning message is displayed if you click a calibration button without waiting for 20 minutes after the instrument powered on. Click the Cancel button and allow at least 20 minutes for the instrument to warm up. AWG5000 Series Service Manual...
  • Page 37: Figure 3-2: 10 Mhz Reference Clock Calibration Test Hookup

    Figure 3-2: 10 MHz Reference clock calibration test hookup 2. Set the frequency counter settings as indicated below: Input impedance ..AC, 50 Ω Gate time ... 0.1 s AWG5000 Series Service Manual...
  • Page 38: Figure 3-3: 10 Mhz Reference Calibration Dialog Box

    Inter-Channel Skew This is a self calibration. This calibration is required to ensure that the inter-channel Calibration skews are within the specified range. This completes the adjustment procedures. Disconnect the equipment from the instrument. AWG5000 Series Service Manual...
  • Page 39 Maintenance...
  • Page 41: Preparation

    Static discharge can damage any semiconductor component in this CAUTION. instrument. Minimize handling of static-sensitive modules. Transport and store static-sensitive modules in their static protected containers or on a metal rail. Label any package that contains static-sensitive modules. AWG5000 Series Service Manual...
  • Page 42: Inspection And Cleaning

    The cabinet helps keep dust out of the instrument and must be in place during normal operation. To avoid damage to the instrument, do not expose it to any sprays, liquids, or solvents. AWG5000 Series Service Manual...
  • Page 43: Table 4-1: External Inspection Checklist

    Avoid the use of chemical cleaning agents, which might damage the CAUTION. plastics used in this AWG5000 Series Arbitrary Waveform Generator. Use only deionized water when cleaning the front-panel buttons. Use an ethyl alcohol solution as a cleaner and rinse with deionized water.
  • Page 44 Inspect the internal portions of the instrument for damage and wear, using Table 4-2 as a guide. Defects should be repaired immediately. To prevent damage from electrical arcing, ensure that circuit boards CAUTION. and components are dry before applying power to the instrument. AWG5000 Series Service Manual...
  • Page 45: Table 4-2: Internal Inspection Checklist

    NOTE. If steps 1 and 2 do not remove all the dust or dirt, please contact Tektronix. (See Contacting Tektronix at the front of this manual.) Lubrication. There is no periodic lubrication required for the AWG5000 Series Arbitrary Waveform Generator.
  • Page 46 Maintenance AWG5000 Series Service Manual...
  • Page 47: Removal And Installation Procedures

    Refer to the Adjustment Procedures section. List of Modules The Replaceable Parts section provides a list of all replaceable modules. Any electrical or mechanical module, assembly, or part listed in the parts list is referred to as a module. AWG5000 Series Service Manual...
  • Page 48: Figure 4-1: Disassembly Procedures For External Modules And Internal

    MIO board Processor module* * To remove the Prosessor module, you need to remove the AWG1G board. Secondary power supply Power supply RFI filter Power board Figure 4-1: Disassembly procedures for external modules and internal modules (top) AWG5000 Series Service Manual...
  • Page 49: Figure 4-2: Disassembly Procedures For Internal Modules (Bottom)

    Screwdriver for removing # 2 size Standard tool Phillips screws 3/16 inch open-end wrench Used to remove nut posts Standard tool Angle-Tip Tweezers Used to remove the front-panel knob Standard tool Wire cutter Used to remove cable ties Standard tool AWG5000 Series Service Manual...
  • Page 50: Procedures For External Modules

    2. Orient the instrument so its top is down on the work surface and its rear is facing you. 3. Remove the two screws securing the cosmetic cover to the chassis. 4. Loosen the cover with a detaching tool, such as a flat-blade screwdriver. 4-10 AWG5000 Series Service Manual...
  • Page 51: Figure 4-3: Handle, Snaps, Cosmetic Covers, And Front-Trim Unit

    5. Tilt the rear side of the cover up, and then pull it off the chassis. Installation. To install, reverse this procedure. Snap Cosmetic cover (top) Front-trim unit Cosmetic cover (bottom) Handle base bracket Figure 4-3: Handle, snaps, cosmetic covers, and front-trim unit removal AWG5000 Series Service Manual 4-11...
  • Page 52 Orient the instrument so that the top is on the work surface. b. Remove the 12 screws securing the bottom EMI cover to the chassis. c. Lift the cover off the chassis. Installation. To install, reverse this procedure. 4-12 AWG5000 Series Service Manual...
  • Page 53: Figure 4-4: Emi Covers Removal

    2. Loosen the two thumb screws securing the hard disk assembly to the instrument. 3. Holding the screw heads, pull out the removable hard disk assembly carefully. Installation. To install, reverse this procedure. AWG5000 Series Service Manual 4-13...
  • Page 54: Figure 4-5: Fan Tray Unit Removal

    4. Lift the fan tray unit away from the chassis. 5. Use a #2 Phillips screwdriver to remove the 12 screws securing the fans to the fan bracket. Install. To install, reverse this procedure. Fan bracket Figure 4-5: Fan tray unit removal 4-14 AWG5000 Series Service Manual...
  • Page 55 Remove the two screws securing the touch screen module to the front-panel bracket. b. Disconnect the cable from J200 on the LCDA board. c. Lift the touch screen module away from the front-panel bracket. AWG5000 Series Service Manual 4-15...
  • Page 56: Figure 4-6: Disassembly Of Front-Panel Assembly

    5. Disassemble the front-panel assembly components, using Figure 4-6 as a guide. Installation. To install, reverse this procedure. FSW board Front-panel bracket Inverter board LCD module Touch screen module LCDA board Front Panel board Figure 4-6: Disassembly of front-panel assembly 4-16 AWG5000 Series Service Manual...
  • Page 57: Figure 4-7: Location Of The Cable Ties

    Removal. To remove the DVD-ROM drive, refer to Figure 4-8 on page 4-18 and follow these steps: 1. Remove the handle unit, snaps, cosmetic covers, front-trim unit, EMI covers, hard disk assembly, and drive-bay module. AWG5000 Series Service Manual 4-17...
  • Page 58: Figure 4-8: Drive Module Removal

    3. Slide the DVD-ROM drive with bracket out from the drive-bay chassis. 4. Remove the four screws securing the DVD-ROM drive to the bracket. Installation. To install, reverse this procedure. DVD-ROM board DVD-ROM drive DVD-ROM drive Drive-bay chassis DVD-ROM bracket Figure 4-8: Drive module removal 4-18 AWG5000 Series Service Manual...
  • Page 59: Figure 4-9: Removing The Clamp Cover

    2. Orient the instrument so that the bottom is on the work surface and the rear is facing you. 3. Remove the clamp cover from the cable clamp to release the ribbon cable. See Figure 4-9. Multi-Input/Output board clamp cover Figure 4-9: Removing the clamp cover AWG5000 Series Service Manual 4-19...
  • Page 60: Figure 4-10: Multi-Input/Output Board Removal

    1. Remove the handle unit, snaps, cosmetic covers, front-trim unit, EMI covers, drive-bay module, and Multi-Input/Output board. 2. Orient the instrument so that the bottom is on the work surface. 3. Remove the two screws securing the drive-bay bracket to the processor tray. 4-20 AWG5000 Series Service Manual...
  • Page 61 The cable from the Front Connector board at J8G2. The cable from the Front Switch board and Multi-Input/Output board at J8J3. 10. Carefully lift up and slide the processor module away from the chassis. Installation. To install, reverse this procedure. AWG5000 Series Service Manual 4-21...
  • Page 62: Figure 4-11: Processor Board Removal

    Removal and Installation Procedures Connector bracket Processor board Drive-bay bracket Heat sink (includes a fan) Processor tray Figure 4-11: Processor board removal 4-22 AWG5000 Series Service Manual...
  • Page 63 You need a screwdriver with a T-15 Torx tip (items 1 and 2). Removal. To remove the power supply, refer to Figure 4-12 and follow these steps: 1. Remove the handle unit, snaps, cosmetic covers, and EMI covers. AWG5000 Series Service Manual 4-23...
  • Page 64: Figure 4-12: Secondary Power Supply, Power Supply, And Rfi Filter

    6. Pull the RFI filter away from the chassis. Installation. To install, reverse this procedure. Power supply Secondary power supply Power supply bracket Insulator RFI filter Figure 4-12: Secondary power supply, power supply, and RFI filter removal 4-24 AWG5000 Series Service Manual...
  • Page 65: Figure 4-13: Power Board Removal

    J611, J630, and J632 on the Power board. 4. Remove the 10 screws securing the Power board to the chassis. 5. Lift the board away from the chassis. Installation. To install, reverse this procedure. Power board Figure 4-13: Power board removal AWG5000 Series Service Manual 4-25...
  • Page 66: Procedures For Internal Modules (Lower)

    3. Disconnect the cables from J100 and J200 on the Front Connector board. 4. Remove the screw securing the Front Connector board to the chassis. 5. Grasp the board and pull it gently away from the chassis. Installation. To install, reverse this procedure. 4-26 AWG5000 Series Service Manual...
  • Page 67: Figure 4-14: Front Connector Board, Clk1G Module, Awg1G Board, And Out1G Boards Removal (4-Channel Model)

    CLK1G module CH3 & CH4 OUT1G board AWG1G board CH1 & CH2 OUT1G board Cable bracket BNC bracket Front Connector board Figure 4-14: Front Connector board, CLK1G module, AWG1G board, and OUT1G boards removal (4-channel model) AWG5000 Series Service Manual 4-27...
  • Page 68 6. Remove the two screws securing the top shield case and CLK1G board to the chassis. 7. Lift the top shield case away from the CLK1G board. 8. Lift the CLK1G board away from the bottom shield case. Installation. To install, reverse this procedure. 4-28 AWG5000 Series Service Manual...
  • Page 69: Figure 4-15: Disassembly Of The Clk1G Module

    Removal and Installation Procedures Top shield case CLK1G board Bottom shield case Figure 4-15: Disassembly of the CLK1G module AWG5000 Series Service Manual 4-29...
  • Page 70: Figure 4-16: Location Of The Cable Clamps On The Awg1G Board

    The cable at J6030B from the AWG1G board to the CH1 & CH2 OUT1G board (J4000). The cable at J7800I from the AWG1G board to the Channel 1 Mkr 1 connector. The cable at J7810I from the AWG1G board to the Channel 1 Mkr 2 connector. 4-30 AWG5000 Series Service Manual...
  • Page 71 1. Remove the handle unit, snaps, cosmetic covers, front-trim unit, EMI covers, and CLK1G module. 2. Orient the instrument so that the top is on the work surface. 3. Remove the screw securing the cable clamp to the AWG1G board. See Figure 4-17. AWG5000 Series Service Manual 4-31...
  • Page 72: Figure 4-17: Location Of The Cable Clamp On The Awg1G Board

    The cable at J6030D from the AWG1G board to the CH3 & CH4 OUT1G board (J4000). The cable at J7800I from the AWG1G board to the Channel 1 Mkr 1 connector. The cable at J7810I from the AWG1G board to the Channel 1 Mkr 2 connector. 4-32 AWG5000 Series Service Manual...
  • Page 73 1. Remove the handle unit, snaps, cosmetic covers, front-trim unit, EMI covers, and AWG1G board. 2. Orient the instrument so that the top is on the work surface. 3. Use a wire cutter to remove the four cable ties. See Figure 4-18. AWG5000 Series Service Manual 4-33...
  • Page 74: Figure 4-18: Location Of The Cable Ties

    Analog connector. The cable at J3180 from the CH1 & CH2 OUT1G board to the Channel 2 Analog connector. The cable at J4180 from the CH1 & CH2 OUT1G board to the Channel 2 Analog connector. 4-34 AWG5000 Series Service Manual...
  • Page 75: Figure 4-19: Location Of The Cable Ties

    The cable at J5000 from the CH3 & CH4 OUT1G board to the AWG1G board (J1060). The cable at J7000 from the CH3 & CH4 OUT1G board to the CH1 & CH2 OUT1G board (J7000). AWG5000 Series Service Manual 4-35...
  • Page 76 The cable at J3410 from the CH3 & CH4 OUT1G board to the Ch 4 Add Input connector. 6. Remove the nine screws securing the CH3 & CH4 OUT1G board to the chassis. 7. Lift the board away from the chassis. Installation. To install, reverse this procedure. 4-36 AWG5000 Series Service Manual...
  • Page 77: Troubleshooting

    PC troubleshooting and Microsoft Windows operating system skills. Details of PC and Windows operation and service are not in this manual. For assistance, contact your local Tektronix Service Center. Equipment Required You will need a digital multimeter to check voltages of the power supply, secondary power supply, and PWR board, as described on page 4-40 through page 4-43.
  • Page 78: Figure 4-20: Primary Troubleshooting Tree (1)

    Is AWG software Replace HDD unit. Replace the μ-ATX board. running? Front panel keys Replace the front-panel and LEDs OK? board. Touch screen Replace the LCDA board or working? touch screen. Figure 4-20: Primary troubleshooting tree (1) 4-38 AWG5000 Series Service Manual...
  • Page 79: Figure 4-21: Primary Troubleshooting Tree (2)

    Execute functional test described in Performance Verification manual. Replace the AWG1G or Amplitude OK? OUT1G board. Replace the CLK1G board. Frequency OK? Output Replace the AWG1G board. waveform OK? Completed. Figure 4-21: Primary troubleshooting tree (2) AWG5000 Series Service Manual 4-39...
  • Page 80: Figure 4-22: Pwr Board Test Points (J104)

    Table 4-4: Test points voltage (J104) Test point Voltage Allowance Pin1 +3.3 V +3.1 V to +3.5 V Pin 3 +5 V +4.7 V to +5.3 V Pin 5 +12 V +11.4 V to +12.6 V 4-40 AWG5000 Series Service Manual...
  • Page 81: Figure 4-23: Secondary Power Supply Test Points (Cn51 And Cn61)

    Table 4-5: Test points voltage (CN51 and CN61) Test point Voltage Allowance CN51: Pin 1 +5 V +4.7 V to +5.3 V CN61: Pin 1 +15 V +14.2 V to +15.8 V CN61: Pin 3 -15 V -14.2 V to -15.8 V AWG5000 Series Service Manual 4-41...
  • Page 82: Figure 4-24: Awg1G Board Test Points (J2000)

    +3.1 V to +3.5 V Pin 12 –5.5 V –5.2 to –5.8 V Pin 13 +15 V +14.2 V to +15.8 V Pin 17 +5 V +4.7 V to +5.3 V Pin 24 –12 V –11.4 to –12.6 V 4-42 AWG5000 Series Service Manual...
  • Page 83: Figure 4-25: Out1G Board Test Points (J7000)

    Pin 2 +8.2 V +7.8 V to +8.6 V Pin 3 +3.3 V +3.1 to +3.5 V Pin 7 -8.2 V -7.8 V to -8.6 V Pin 8 -15 V +14.2 V to +15.8 V AWG5000 Series Service Manual 4-43...
  • Page 84: Figure 4-26: Diagnostics Dialog Box

    To run the instrument Diagnostics manually, do these steps: 1. Turn off all other applications. 2. From the menu bar, select System and then select Diagnostics...; the Diagnostics dialog box appears (see Figure 4-26). Figure 4-26: Diagnostics dialog box 4-44 AWG5000 Series Service Manual...
  • Page 85: Table 4-8: Diagnostics Error

    Fail expected data=xxxx actual data=xxxx TNO: TEK0015 number REG: Register number 120000023 HwDiagAwgRegister AD9736 failed channel=1 address=xx AWG1G AWG1G: U1000, U6000A expected data=0xXX actual data=0xXX 120000024 HwDiagAwgRegister AD9736 failed channel=2 address=xx AWG1G AWG1G: U1000, U6000B expected data=0xXX actual data=0xXX AWG5000 Series Service Manual 4-45...
  • Page 86 IC: IC number Pin: Pin number 122000002 HwDiagWaveformMemory channe2, Memory Cell testNo AWG1G AWG1G: U4000B, U4500B, U4510B, BIT[CNT] Error U4520B, U4530B, U4600B, U4610B, testNo: 1'st or 2'nd U4620B, U4630B. BIT: Data bus bit CNT: data bit 4-46 AWG5000 Series Service Manual...
  • Page 87 CLK1G: U3200 125000002 HwDiagDataTiming channel=3, Value is out of a range AWG1G, CLK1G AWG1G: U1000 max=xx min=xx CLK1G: U3200 125000003 HwDiagDataTiming channel=4, Value is out of a range AWG1G, CLK1G AWG1G: U1000 max=xx min=xx CLK1G: U3200 AWG5000 Series Service Manual 4-47...
  • Page 88 Analog Level Diag CHX positive (negative) 12db attenuator OUT1G OUT1G: K1105, K2105, K3105, K4105 failed. Expected 0.2530, actual xxxxx 141XP0033 Analog Level Diag CHX positive (negative) 24db attenuator OUT1G OUT1G: K1110, K2110, K3110, K4110 failed. Expected 0.0631, actual xxxxx 4-48 AWG5000 Series Service Manual...
  • Page 89 Analog Level Diag CHX positive (negative) 6db attenuator OUT1G OUT1G: K1100, K2100, K3100, K4100 failed. Expected 0.5005, actual xxxxx 151XP0032 Analog Level Diag CHX positive (negative) 12db attenuator OUT1G OUT1G: K1105, K2105, K3105, K4105 failed. Expected 0.2530, actual xxxxx AWG5000 Series Service Manual 4-49...
  • Page 90 Analog Level Diag CHX amp gain balance failed. OUT1G OUT1G: U1140, U1145, U1160A/B, U2140, U2145, U2160A/B, U3140, U3145, U3160A/B, U4140, U4145, U4160A/B 151XP0058 Analog Level Diag CHX A/D converter read failed. OUT1G OUT1G: U5010, U5100, U5112, U5114 4-50 AWG5000 Series Service Manual...
  • Page 91: Figure 4-27: Calibration Dialog Box

    U3145, U3160A/B, U4140, U4145, U4160A/B 200XP0037 Analog Level Calibration CHX positive (negative) output offset OUT1G OUT1G: U1200, U1220, Q1218, 0V failed. Expected:0.0V, actual:xxxV Q1228, U2200, U2220, Q2218, Q2228, U3200, U3220, Q3218, Q3228, U4200, U4220, Q4218, Q4228 AWG5000 Series Service Manual 4-51...
  • Page 92 Analog Level Calibration CHX filter 100mhz gain balance failed. OUT1G OUT1G: K1120, K1125, K2120, K2125, K3120, K3125, K4120, K4125 200X00057 Analog Level Calibration CHX amp gain balance failed. OUT1G OUT1G: U1140, U1145, U1160A/B, U2140, U2145, U2160A/B, U3140, U3145, U3160A/B, U4140, U4145, U4160A/B 4-52 AWG5000 Series Service Manual...
  • Page 93: Table 4-10: Hardware Error

    AWG1G AWG1G: U1000, U4310B, U4420B 300000060 Failed in the phase adjustment of two Tek0015. channel=3 AWG1G AWG1G: U1000, U4310C, U4420C 300000061 Failed in the phase adjustment of two Tek0015. channel=4 AWG1G AWG1G: U1000, U4310D, U4420D AWG5000 Series Service Manual 4-53...
  • Page 94 MIO, AWG1G MIO: U300, AWG1G: U1000, U4000x Product configuration (number of channels) is wrong. 300000214 The download (AwgController-DONE) timeout period has expired. MIO,AWG1G MIO: U300, AWG1G: U1000, U4000x Product configuration (number of channels) is wrong. 4-54 AWG5000 Series Service Manual...
  • Page 95: After Repair

    After removal and replacement of a module, you must perform the Adjustment and Performance Verification procedures. You can find the information about these procedures in the following manuals: AWG5000 Series Arbitrary Waveform Generators Service Manual (Tektronix part number 071-2083-XX) AWG5000 Series Arbitrary Waveform Generators Specifications and...
  • Page 96: Returning The Instrument For Service

    Reason for returning. A complete description of the required service. Mark the address of the Tektronix Service Center and the return address on the shipping carton in two prominent locations. See www.tektronix.com\service to find a Tektronix Service Center near you.
  • Page 97 Replaceable Parts...
  • Page 99: Parts Ordering Information

    Replaceable Parts This section contains a list of the replaceable modules for the AWG5000 Series Arbitrary Waveform Generators. Use this list to identify and order replacement parts. Parts Ordering Information Replacement parts are available through your local Tektronix field office or representative.
  • Page 100 Replaceable Parts Part Number Revision Tektronix part numbers contain two digits that show the revision level of the part. Level For most parts in this manual, you will find the letters XX in place of the revision level number. Part Number Revision Level...
  • Page 101: Using The Replaceable Parts List

    Using the Replaceable Parts List This section contains a list of the mechanical and/or electrical components that are replaceable for the AWG5000 Series. Use this list to identify and order replacement parts. The following table describes each column in the parts list.
  • Page 102 MARKER, IDENTIFICATION, FRONT PANEL, 4 CHANNEL, SAFETY CONTROLLED 335-1579-00 MARKER, IDENTIFICATION, SIGNAL INPUT/OUTPUT, BNC, 4 CHANNEL, SAFETY CONTROLLED 335-1580-00 MARKER, IDENTIFICATION, SIGNAL INPUT/OUTPUT, BNC, 2 CHANNEL, SAFETY CONTROLLED 366-0841-00 KNOB, CAP; 1.380 DIA, PC/ABS, SILVER GRAY AWG5000 Series Service Manual...
  • Page 103: Figure 5-1: Exploded Diagram - Cosmetics

    Replaceable Parts Figure 5-1: Exploded diagram - Cosmetics AWG5000 Series Service Manual...
  • Page 104: Figure 5-2: Exploded Diagram - Emi Covers

    Serial no. Serial no. number number effective dis-cont’d Name & description EMI COVERS 200-4992-00 COVER, EMI, TOP 200-4993-00 COVER, EMI, BOTTOM 211-1050-00 SCREW,MACHINE; 6-32 X 0.312 L,PNH,STL CAD PLT,T15 Figure 5-2: Exploded diagram - EMI Covers AWG5000 Series Service Manual...
  • Page 105: Figure 5-3: Exploded Diagram - Chassis Assembly

    Serial no. number number effective dis-cont’d Name & description CHASSIS ASSEMBLY 131-6417-00 CONTACT,ELEC; GROUNDING,0.600 L X 0.250 W X 0.220 D,ELECTROLESS NICKEL PLATE 348-0966-00 GROMMET,PLASTIC;BLACK,ROUND,0.625 ID 441-2471-00 CHASSIS, MAIN; ASSEMBLY Figure 5-3: Exploded diagram - Chassis assembly AWG5000 Series Service Manual...
  • Page 106 HOLES: 156 IN. /.187 IN. TOP/BOTTOM PANEL THK: .062 IN.; 671-6063-00 CIRCUIT BOARD ASSY;FRONTPANEL 389377400 WIRED 260-2819-00 SWITCH,KEYPAD; FRONT PANEL 4CH; AWG7000 SERIES 407-5176-00 BRACKET;TRIM,FRONTPANEL,SHEET METAL,SAFETY CONTROLLED 174-4791-00 CA ASSY; DISPLAY ADAPTER TO XGA DISPLAY,SAFETY CONTROLLED 211-1050-00 SCREW,MACHINE; 6-32 X 0.312 L,PNH,STL CAD PLT,T15 AWG5000 Series Service Manual...
  • Page 107: Figure 5-4: Exploded Diagram - Front-Panel Assembly

    Replaceable Parts (includes 4) Figure 5-4: Exploded diagram - Front-panel assembly AWG5000 Series Service Manual...
  • Page 108 BRACKET, ASSEMBLY, REMOVABLE 3.5 INCH HARD DRIVE CARTRIDGE, AWG5000 SERIES 211-0734-00 SCREW,MACHINE; 6-32 X 0.250,FLH100,STL,CDPL,T-10 TORX DR 174-5017-00 CABLE ASSEMBLY; DVD POWER, SAFETY CONTROLLED 065-0768-00 ACCESSORY KIT; REMOVABLE HDD ASSY; UNPROGRAMMED HARD DRIVE 119718900 W/PACKAGING 5-10 AWG5000 Series Service Manual...
  • Page 109: Figure 5-5: Exploded Diagram - Drive Module

    Replaceable Parts Figure 5-5: Exploded diagram - Drive module AWG5000 Series Service Manual 5-11...
  • Page 110 407-5205-00 BRACKET,CONN; AL,0.063 THICK,W/131641700 EMI CLIPS 211-1050-00 SCREW,MACHINE; 6-32 X 0.312 L,PNH,STL CAD PLT,T15 214-3903-00 SCREW,JACK; 4-40 X 0.312 LONG,0.188 H HEX HEAD STAND OFF,4-40 INT THD, X 0.312 THD EXT 4-40 065-0761-00 MODULE; PROCESSOR 5-12 AWG5000 Series Service Manual...
  • Page 111: Figure 5-6: Exploded Diagram - Processor Module

    Replaceable Parts Figure 5-6: Exploded diagram - Processor module AWG5000 Series Service Manual 5-13...
  • Page 112 SCREW,MACHINE; 6-32 X 0.312 L,PNH,STL CAD PLT,T15 119-B128-00 POWER SUPPLY;ZWQ130-5224 407-5145-00 BRACKET, POWER SUPPLY 211-1050-00 SCREW,MACHINE; 6-32 X 0.312 L,PNH,STL CAD PLT,T15 174-5269-00 CABLE ASSY; AC INLET TO CKT BD,14 IN L,AWG16 X 3,W/FN9222-10/06 FILTER/1-480700 CONN 5-14 AWG5000 Series Service Manual...
  • Page 113: Figure 5-7: Exploded Diagram - Power Supply Assembly

    Replaceable Parts Figure 5-7: Exploded diagram - Power supply assembly AWG5000 Series Service Manual 5-15...
  • Page 114 671-6049-01 CIRCUIT BOARD ASSY; MIO MULTI-I/O, FR-4 343-1627-00 CABLE,CLAMP; 0.5H X 2.98L,0.625 W,NYLON 6-6,ADHESIVE,COVER 343-1636-00 343-1636-00 CLAMP, COVER; FLT CABLE ROUTING CLAMP, 343-1627-00 BASE 335-1723-00 MARKER,IDENT; LABEL,REAR IO,AWG2,SAFETY CONTROLLED,LEXAN,4CH 335-1662-00 MARKER,IDENT; LABEL,REAR I/O,MAIN,LEXAN, SAFETY CONTROLLED 5-16 AWG5000 Series Service Manual...
  • Page 115: Figure 5-8: Exploded Diagram - Pci/Power Interface Module

    Replaceable Parts (includes 5) (includes 20 and 21) Figure 5-8: Exploded diagram - PCI/power interface module AWG5000 Series Service Manual 5-17...
  • Page 116: Figure 5-9: Exploded Diagram - Fan Tray Assembly

    211-1213-01 SCREW, METRIC 5 X 10MM, FLATHEAD, STL, ZINC, POSI 407-5170-00 BRACKET,FAN TRAY 343-0549-00 STRAP,TIEDOWN; 0.098 W X 4.0 L, ZYTEL 119-7182-00 FAN,TUBEAXIAL; 12VDC,0.40A,4.80W,2500RPM,100.6CFM,120MM,LEADS W/CONN; SAFETY CONTROLLED Figure 5-9: Exploded diagram - Fan tray assembly 5-18 AWG5000 Series Service Manual...
  • Page 117: Figure 5-10: Exploded Diagram - Usb Module

    CA ASSY,INTCON; CMPST,USB2.0,24/28 AWG,24 IN L,10 POS,1 X 10 AND 2 X 5,USB2.0 CABLE X 2,0.1 CTR 174-5248-00 CABLE ASSY ELEC: 1 X 8P,FLAT,0.1,16.8IN L,AWG 26,103958-7 AMP 211-1050-00 SCREW,MACHINE; 6-32 X 0.312 L,PNH,STL CAD PLT,T15 (includes 1 and 2) Figure 5-10: Exploded diagram - USB module AWG5000 Series Service Manual 5-19...
  • Page 118 SCREW,MACHINE; 6-32 X 0.312 L,PNH,STL CAD PLT,T15 211-1050-00 SCREW,MACHINE; 6-32 X 0.312 L,PNH,STL CAD PLT,T15 174-5303-00 CABLE ASSY ELEC: 2X4, 9.5IN L,18 AWG X 8, 39-01-2085, MOLEX 211-1050-00 SCREW,MACHINE; 6-32 X 0.312 L,PNH,STL CAD PLT,T15 5-20 AWG5000 Series Service Manual...
  • Page 119: Figure 5-11: Exploded Diagram - Awg1G/Clk1G/Out1G Modules

    Replaceable Parts CLK1G module See Figure 5-12 on page 5-22 for detailed view. (includes 4) AWG1G board See Figure 5-13 on page 5-23 for detailed view. Figure 5-11: Exploded diagram - AWG1G/CLK1G/OUT1G modules AWG5000 Series Service Manual 5-21...
  • Page 120: Figure 5-12: Exploded Diagram - Clk1G Module

    SCR, ASSEM WSHR; 6-32 X 0.625, PNH, 410 SS PASSIVATED, T-15 TORX DR, MACHINE 337-4376-00 SHIELD ELEC,TOP,CLK1 BOARD 671-6187-00 CLK1G, 1.2GHZ CLOCK BOARD ASSY 337-4377-00 SHIELD ELEC,BOTTOM,CLK1 BOARD Figure 5-12: Exploded diagram - CLK1G module 5-22 AWG5000 Series Service Manual...
  • Page 121: Figure 5-13: Exploded Diagram - Awg1G Module

    CIRCUIT BOARD ASSY; AWG1G, AWG 1.2GS, 4CH, W/EXT_MEMORY 174-5316-00 AWG50x2 CABLE ASSY ELEC: 1 X 5,7.1IN L,24 AWG X 5,PHR-5,PHR-5,JST 174-5316-00 AWG50x4 CABLE ASSY ELEC: 1 X 5,7.1IN L,24 AWG X 5,PHR-5,PHR-5,JST Figure 5-13: Exploded diagram - AWG1G module AWG5000 Series Service Manual 5-23...
  • Page 122 CABLE ASSY ELEC: RF,COAX,9IN L,50 OHM,PELTOLA,FLEX COAX CABLE 131-8054-00 AWG50x2 CONN,RF,JACK; BNC,PNL,50 OHM,FEMALE,STR,PELTOLA/PNL MNT,SILVER ALLOY,0.576 MLG X 0.366 TERMN,0.375-32 THD,D-FLAT, WITH KNURL 131-8054-00 AWG50x4 CONN,RF,JACK; BNC,PNL,50 OHM,FEMALE,STR,PELTOLA/PNL MNT,SILVER ALLOY,0.576 MLG X 0.366 TERMN,0.375-32 THD,D-FLAT, WITH KNURL 5-24 AWG5000 Series Service Manual...
  • Page 123: Figure 5-14: Exploded Diagram - Bnc Bracket

    Replaceable Parts Figure 5-14: Exploded diagram - BNC bracket AWG5000 Series Service Manual 5-25...
  • Page 124: Figure 5-15: Exploded Diagram - Rear Connectors

    CABLE ASSY ELEC: RF,COAX,12.8IN L,50 OHM,PELTOLA,FLEX COAX CABLE 131-8054-00 CONN,RF,JACK; BNC,PNL,50 OHM,FEMALE,STR,PELTOLA/PNL MNT,SILVER ALLOY,0.576 MLG X 0.366 TERMN,0.375-32 THD,D-FLAT, WITH KNURL 174-5315-00 CABLE ASSY ELEC: RF,COAX,7.1IN L,50 OHM,PELTOLA,FLEX COAX CABLE Figure 5-15: Exploded diagram - Rear connectors 5-26 AWG5000 Series Service Manual...