Amplitude And Offset Accuracy Tests (Normal Out) - Tektronix AWG610 User Manual

Arbitrary waveform generator
Table of Contents

Advertisement

Appendix B: Performance Verification

Amplitude and Offset Accuracy Tests (Normal Out)

Check Amplitude
Accuracy
B-24
These procedures check the accuracy of the amplitude and offset outputs of the
AWG610 Arbitrary Waveform Generator.
NOTE. The amplitude and offset accuracy checks are structured as a continuous
test. The next test uses the control settings from the previous test and uses the
next step in the sequence file.
NOTE. When you output signal from the CH1 or CH1 OUTPUT, check that the
other OUTPUT ( CH1 or CH1 ) LED is off.
If the other OUTPUT LED is on, push the CH1 or CH1 OUT button to turn off
the output.
Equipment
A 50 W BNC coaxial cable, a 50 W precision terminator,
required
two
adapter, and a digital multimeter (DMM).
Prerequisites
The AWG610 Arbitrary Waveform Generator must meet the
prerequisites listed on page B-8.
Do the following steps to install the test hookup and set the test equipment
controls:
1. Use a 50 W BNC coaxial cable, an SMA(Fe)-BNC(Ma) adapter, a 50 W
precision terminator, and a BNC-to-dual banana adapter to connect the
AWG610 Arbitrary Waveform Generator CH1 output to the DMM input
connector (see Figure B–9).
SMA(Ma)-BNC(Fe) adapters,
AWG610 Arbitrary Waveform Generator User Manual
BNC (female) to dual banana

Advertisement

Table of Contents
loading