Tektronix AWG5200 Series Technical Reference page 45

Arbitrary waveform generators, specifications and performance verification
Hide thumbs Also See for AWG5200 Series:
Table of Contents

Advertisement

Checking the marker
outputs
AWG5200 Series Technical Reference
Required equipment
Oscilloscope
Four TCA-SMA adapters
Four 50 Ω SMA cables
1. Press the All Outputs Off button on the generator (toggle from green to red)
to disable all outputs.
2. Connect the generator's channel 1 markers to the test oscilloscope using a
50 Ω SMA cable and a TCA-SMA adapter.
Connect CH1:1 to channel 1 of the test oscilloscope.
Connect marker CH1:2 to channel 2 of the test oscilloscope.
Connect marker CH1:3 to channel 3 of the test oscilloscope.
Connect marker CH1:4 to channel 4 of the test oscilloscope.
NOTE.
If a channel's marker is not connected to the test oscilloscope, it must
be terminated with a 50 Ω SMA terminator.
3. Set the test oscilloscope as follows:
a. Vertical scale: 1 V/div (CH 1 through CH 4)
b. Horizontal scale: 100 ns/div
c. Input coupling: DC
d. Input impedance: 50 Ω
Brief procedures
Prerequisites
None
33